The OPSA-150 is a new organic pollutant monitor that uses HORIBA's proprietary Rotary Cell Length Modulation, a measuring technique incorporating 25 years of expertise. The unit can be used as an organic pollutant monitor at drainage systems for determining compliance with COD/BOD monitoring regulation, for monitoring quality of water measuring levels of organic matter at water supply intakes, ands as an organic monitor on process lines (phenol meter).
Sensor that makes full use of Rotary Cell Length Modulation (Patented)
Zero drift correction is performed in every cycle of measurement. This correction can eliminate the effect of any interference on cells.
Cell length modulation provides readings from multiple measurements, making available results from various cell lengths. This allows the single device to make measurements on a large range of concentrations from as low as 0 to 0.1 Abs, up to 0 to 5.0 Abs.
MODEL NAME: | OPM300-OPM200 |
Measurement Methods | UV + VIS Spectrophotometer (Dual Beam AbsorptionSpectrophotometry) 200-350nm for Ultraviolet spectrum scanning 320-750nm for Visible spectrum scanning* |
Measurement items | COD, BODc, TSS, TOCc |
Measurement Range | COD (0-300 mg/l), BODc (0-300 mg/l), TSS (0-300 mg/l), TOCc (0-100mg/l) |
Measurement Accuracy | OPM300 Within ± 5% of full scale, OPM200 Within ± 7% of full scale |
Measurement Repeat | OPM300 - OPM200 Within ± 5% of full scale |
Measurement Linearity | OPM300 - OPM200 Within ± 5% of full scale |
Measurement Stability | OPM300 - OPM200 Within ± 2% of full scale |
Measurement Interval | T90 in One Second, Data output within 1 minute |
Resolution | 0.01mg/L |
Calibration | Multi point based automatic calibration. Standard calibration as per factory setting. Local calibration for adaptive environments. For COD compensation with respect to organics and inorganic proportions. |
Cable length | 10 m standard (Expandable up to 30m) |
Sample minimum pressure | 1 Bar |
Power | 230VAC±10%, 50Hz |
Output | RS485 Smart Two way Communication Enabled |
Protection class | IP68 |
Cleaning Method | Advanced Surface conductive cleaning |
MOC | OPM300 Titanium Quartz POM SS316L, OPM200 Quartz POM SS316L |
Sensor Diagnostics | Onboard library for diagnostic commands available |
Display | Anti-glare LCD Display – Touch Panel Type |
Ambient operating Condition | Ambient Temperature 0-55 ̊C, Ambient Humidity 90% RH |
Storage temperature | 5~45°C |
Background Data Input | Sample loss input, Time correction input, Flux Cancellation Input, Temperature Compensation Input |
Construction | For Outdoor Installation |
Color/Finish | Standard material finish as per MOC |
Installation Conditions | *Installation guide book to be referred |
MODEL NAME: | OPM250 |
Measurement Methods | UV + VIS Spectrophotometer (Dual Beam AbsorptionSpectrophotometry) 180-350nm for Ultraviolet spectrum scanning 320-750nm for Visible spectrum scanning* |
Measurement items | COD, BODc, TSS, TOCc |
Measurement Range | COD (0-500-3500 mg/l), BODc (0-500-3500 mg/l), TSS (0-500-3500 mg/l), TOCc (0-500-2000 mg/l) |
Measurement Accuracy | Within ± 8% of full scale |
Measurement Repeat - Ability | Within ± 8% of full scale |
Measurement Linearity | Within ± 8% of full scale |
Measurement Stability | Within ± 5% of full scale |
Measurement Interval | T90 Data output within 1 minute |
Resolution | 0.01mg/L |
Calibration | Multi point based automatic calibration. Standard calibration as per factory setting. Local calibration for adaptive environments. For COD compensation with respect to organics and inorganic proportions. |
Cable length | 10 m standard (Expandable up to 30m) |
Sample minimum pressure | 1 Bar |
Power | 230VAC±10%, 50Hz |
Output | RS485 Smart Two way Communication Enabled |
Photo incident Source | UV + Vis entire spectral Light incidence |
Protection class | Ingress Protection IP68, EMC Conformity with EN61323-1, EN61010-1 |
Cleaning Method | Advanced Surface conductive cleaning / Compressed Air Cleaning |
MOC | Titanium Quartz POM SS316L |
Sensor Diagnostics | Onboard library for diagnostic commands available |
Display | Anti-glare LCD Display – Touch Panel Type |
Ambient operating Condition | Ambient Temperature 0-55 ̊C, Ambient Humidity 90% RH |
Storage temperature | 5~45°C |
Background Data Input | Sample loss input, Time correction input, Flux Cancellation Input, Temperature Compensation Input |
Construction | For Outdoor Installation |
Color/Finish | Standard Integral material finish as per MOC |
Installation Conditions | *Installation guide book to be referred |
MODEL NAME: | OPM3000 |
Measurement Methods | UV + VIS Spectrophotometer (Dual Beam AbsorptionSpectrophotometry) 180-350nm for Ultraviolet spectrum scanning 320-750nm for Visible spectrum scanning* |
Measurement items | COD, BODc, TSS, TOCc |
Measurement Range | COD (0-10000-30000 mg/l), BODc (0-10000-30000 mg/l), TSS (0-10000-30000 mg/l), TOCc (0-5000-10000 mg/l) |
Measurement Accuracy | Within ± 2% of full scale |
Measurement Repeat - Ability | Within ± 5% of full scale |
Measurement Linearity | Within ± 5% of full scale |
Measurement Stability | Within ± 2% of full scale |
Measurement Interval | T90 Data output within 1 minute |
Resolution | 0.01mg/L |
Calibration | Multi point based automatic calibration. Standard calibration as per factory setting. Local calibration for adaptive environments. For COD compensation with respect to organics and inorganic proportions. |
Cable length | 10 m standard (Expandable up to 30m) |
Sample minimum pressure | 1 Bar |
Power | 230VAC±10%, 50Hz |
Output | RS485 Smart Two way Communication Enabled |
Photo incident Source | UV + Vis entire spectral Light incidence |
Protection class | Ingress Protection IP68, EMC Conformity with EN61323-1, EN61010-1 |
Cleaning Method | Advanced Surface conductive cleaning / Compressed Air Cleaning |
MOC | Titanium Quartz POM SS316L |
Sensor Diagnostics | Onboard library for diagnostic commands available |
Display | Anti-glare LCD Display – Touch Panel Type |
Ambient operating Condition | Ambient Temperature 0-55 ̊C, Ambient Humidity 90% RH |
Storage temperature | 5~45°C |
Background Data Input | Sample loss input, Time correction input, Flux Cancellation Input, Temperature Compensation Input |
Construction | For Outdoor Installation |
Color/Finish | Standard Integral material finish as per MOC |
Installation Conditions | *Installation guide book to be referred |